Loading

SCR121

Code: SCR121
  • resistance temperature sensor
  • temperature range -50...550°C
  • operating temperature of connection heads max. 150°C
  • stainless steel sheath
  • sensor with a replaceable measuring insert
  • possibility of mounting a 4…20 mA or 0…10 V measuring transmitter
  • connection head DANW with local display
Description

The resistance thermometer SCR121 consists of an exchangeable measuring insert, outer protective tube (thermowell) with neck, and aluminum connection head. Intended for measuring the temperature of pipelines, tanks, boilers in difficult conditions. Mounting a temperature transmitter with 4...20 mA or 0...10V output signal is possible.

The measuring insert represents the replaceable element of the complete sensor, which reduces the time and costs of maintenance of the measuring apparatus installed in the object. Spring fixation of the measuring insert provides perfect pressure to the bottom of the protecting tube, reduces the time of reaction of temperature changes, and increases the accuracy. It also reduces natural vibration. Thus, mechanical and electrical defects can be avoided.

Application areas:

  • technological installations in various industries,
  • temperature measurement of pipelines, tanks, boilers,
  • measurement of all media (gases, liquids, solids).

Specification
Measurement range
-50...+550°C
Measuring element
Pt100 resistor
Pt500 resistor
Pt1000 resistor
Accuracy
A class
B class
Sheath
stainless steel 1.4541
diameter 24 mm
1.7335 steel
diameter 18 mm
threaded drilled thermowell
Connection head
B, IP65, -40...150ºC, aluminum
NA, IP65, -40...150ºC, aluminum
DAN, IP65, -40...150ºC, aluminum
DANW, IP65, -40...150ºC, aluminum
Files
Data sheet